PROFIBUS Interface for ultra-wave
TM
, Sonologic II & Weigh II
- 3 -
The ultra‑wave™ /Sonologic II/Weigh II Probus‑DP slave includes a high‑speed interface card using the Siemens
SPC3 chip set and RS485 transmission technology. The bus structure permits the addition and removal of
stations or step‑by‑step commissioning of the system. The card has the following capabilities and requirements:
ultra‑wave™ Compatibility EEPROM must be Rev. P or higher.
SonoII Compatibility EEPROM must be Rev. P or higher.
Transmission Speed Between 9.6 kbit/sec. and 12 Mbit/sec. The ultra‑wave™/ Sonologic II/Weigh II
Probus DP card will auto‑select the speed of the bus.
Network Topology Linear bus, active bus termination on both ends, stub lines only permitted for baud
rates of ≤ 1.5 Mbit/sec. (Termination not supplied)
Medium
Shielded twisted‑pair cable. Siemens 6XV1‑830‑OAH10, Belden 3079A or equivalent
Number of Stations 126 stations on one bus.
Connector
9‑pin D‑sub plug connector. Siemens Bus Connector 6ES7‑972‑OBA11‑OXAO or equivalent
See FIGURE 1. Seimens Bus Connector
Cable Pin Denition D‑sub shell – Ground (outerbraided shield)
Pin 3 – Signal “B1/B2” (Red)
Pin 4 – RTS Signal
Pin 5 – Data ground
Pin 6 – 5VDC
Pin 8 – Signal “A1/A2” (Green)
BAUDE RATES 9.6k‑167.5k Bd 500k Bd 1.5m Bd 3‑12m Bd
Per segment 1000m/3937ft 400m/1312ft 200m/656ft 100m/328ft
With repeater 10,000m/3937ft 400m/1312ft 200m/656ft 100m/328ft
Per ber optic segment 15km/9 mi 15km/9 mi 15km/9 mi 15km/9 mi
Total with ber optic >200km/124 mi >200km/124 mi >200km/124 mi 130km/80 mi
FIGURE 1. Seimens Bus Connector