Vue de la page 190
6-9
1010NFM-3J
Section 6
6.5 SPECIFICATIONS – KYNAR PVDF FLOW TUBE
*Typical values, subject to change without notice.
6.4 SPECIFICATIONS – CPVC FLOW TUBE
*Typical values, subject to change without notice.
Pressure Drop
0.28 psi @ 1.5 GPM (5.7 LPM) 0.24 psi @ 4.5 GPM (17.0 LPM)
flowing water at 68°F flowing water at 68°F
Size 1 Size 2
Liquid Must be compatible with CPVC under Must be compatible with CPVC under
operating conditions and free of sus- operating conditions and free of sus-
pended gasses or solids. pended gasses or solids.
Sensitivity 0.0005 GPM (0.002 LPM) 0.0015 GPM (0.0057 LPM)
Linearity 1% of flow for rates above 0.09 GPM 1% of flow for rates above 0.27 GPM
(0.34 LPM) (1.0 LPM)
Repeatability ±0.2% of flow for rates above 0.6 GPM ±0.2% of flow for rates above 1.5 GPM
(2.3 LPM) (5.7 LPM)
Zero Stability ±0.0013 GPM (0.005 LPM) ±0.003 GPM (0.011 LPM)
Accuracy, ±1% of flow over 0.15 GPM (0.57 ±1% of flow over 0.5 GPM (1.9 LPM) to
Intrinsic LPM) to 5 GPM (20 LPM) 25 GPM (95 LPM)
Accuracy, Flow ±0.5% of flow over 0.15 GPM (0.5 ±0.5% of flow over 0.5 GPM (1.9 LPM)
Calibrated LPM) to 5 GPM (20 LPM) to 25 GPM (95 LPM)
Connection
3
/
8
” FPT ¾” FPT
Repeatability ±0.2% of flow for rates above 0.6 ±0.2% of flow for rates above 1.5 GPM
GPM (2.3 LPM) (5.7 LPM)
Accuracy, ±1% of flow over 0.15 GPM (0.57 ±1% of flow over 0.5 GPM (1.9 LPM) to
LPM) to 5 GPM (20 LPM) 5 GPM (95 LPM)
Zero Stability ±0.0013 GPM (0.005 LPM) ±0.003 GPM (0.011 LPM)
Accuracy, Flow ±0.5% of flow over 0.15 GPM (0.57 ±0.5% of flow over 0.5 GPM (1.9 LPM)
LPM) to 5 GPM (20 LPM) to 25 GPM (95 LPM)
Size 1 Size 2
Liquid M
ust be compatible with PVDF Must be compatible with PVDF under
under operating conditions and free operating conditions and free of
of suspended gasses or solids. suspended gasses or solids.
Pressure Drop 0.28 psi @ 1.5 GPM (5.7 LPM) 0.24 psi @ 4.5 GPM (17.0 LPM)
flowing water at 68°F flowing water at 68°F
Sensitivity 0.0005 GPM (0.002 LPM) 0.0015 GPM (0.006 LPM)
Linearity 1% of flow for rates above 0.09 GPM 1% of flow for rates above 0.27 GPM
(0.34 LPM) (1.0 LPM)
Intrinsic
Calibrated
Connection
3
/
8
” FPT
3
/
4
” FPT