SITRANSRemote DisplaysSITRANS RD100Operating Instructions 01/2011
Page 8 SITRANS RD100 – INSTRUCTION MANUAL 7ML19985JU01SetupFor calibration, a calibrated current source and a screwdriver are required.Calibration con
7ML19985JU01 SITRANS RD100 – INSTRUCTION MANUAL Page 9further details. Wall mounting holes are located in each corner of the enclosure (see Mounting o
www.siemens.com/processautomationPrinted in Canadawww.siemens.com/weighingwww.siemens.com/levelFor more informationwww.siemens.com/processautomationSi
© Siemens AG 2011 Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that of others, and to protect the pro
7ML19985JU01 SITRANS RD100 – INSTRUCTION MANUAL Page 1SITRANS RD100 Loop Powered MeterSITRANS RD100SITRANS RD100 is a 2-wire loop powered, NEMA 4X enc
Page 2 SITRANS RD100 – INSTRUCTION MANUAL 7ML19985JU01SpecificationsDisplay• 1.0 " (25.4 mm) high LCD, numeric display range from -1000 to 1999En
7ML19985JU01 SITRANS RD100 – INSTRUCTION MANUAL Page 3Calibration • Two-step; non-interacting low and highCalibration Range• 4 mA input: display of -1
Page 4 SITRANS RD100 – INSTRUCTION MANUAL 7ML19985JU01MountingWall MountingDimensionsA: 80 mm (3.15") C: 60 mm (2.36") E: 65 mm (2.56")
7ML19985JU01 SITRANS RD100 – INSTRUCTION MANUAL Page 5Optional Panel Mounting KitInstructions1. Insert the RD100 through the mounting plate.2. Mount t
Page 6 SITRANS RD100 – INSTRUCTION MANUAL 7ML19985JU01ConnectionsCalibrator connected to input signal PCBControl loop connected to input signal PCBdis
7ML19985JU01 SITRANS RD100 – INSTRUCTION MANUAL Page 7Calibrator connected to display PCBWARNINGS: • To maintain hazardous area protection, the input